Troubleshooting
Maintenance, Cleaning & Troubleshooting
11-15
Information
Signals
If there is a connection failure within the IntelliVue Clinical Network, an
information signal will be generated. This information signal will be displayed
in the system message area on the Information Center where the affected
wireless patient monitoring device(s) (transceiver, access point, access point
controller, sync unit, or router) is assigned, as well as on all other Information
Centers connected to a common Database Server. The condition causing the
failure will be described in the Wireless Status Log, which is available in
Service Mode.
Information Signals
Information
Signal
Description
What to Do
Wireless
monitoring loss -
Contact Service
Note—
This
information signal
appears on ALL
Information Centers
connected to a
common Database
Server.
Problem with wireless
network infrastructure
device(s).
Contact Service.
Note—
Details about the
communication disruption
are available in the
Wireless Status Log.
Router failure
Call Service
Note—
This
information signal
appears on ALL
Information Centers
connected to a
common Database
Server.
A router used in the
wireless network has
failed.
Contact Service.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...