Telemetry Functions at the Information Center
9-8
Telemetry Functions at the Information Center & TeleMon
Suppress
SpO
2
INOPs
with NBP
Enable/disable the SpO
2
algorithm to suppress sending
technical alarms from the
transceiver during an NBP
measurement for 60 seconds.
Warning
Warning
If you measure SpO
2
on a limb
that has an inflated NBP cuff, a
non-pulsatile SpO
2
technical
alarm can occur. If the monitor
is configured to suppress this
alarm, there can be a delay of
up to 60 seconds in indicating
critical patient status, such as
sudden pulse loss or hypoxia.
enable
disable
enable
Pleth Wave
Enable/disable the transmission
of the Pleth wave (and its
subsequent display) to the
Information Center. For
Continuous SpO
2
mode only.
Note—
When enabled, the Pleth
wave replaces the Vb wave in the
Patient Window during 6-lead
monitoring.
enable
disable
disable
(Pleth is not
displayed.)
Pulse
Enable/disable display of the
Pulse rate at the Information
Center. For Continuous SpO
2
mode only.
enable
disable
disable
(Pulse rate is
not
displayed.)
Patient-Configurable Settings in Telemetry Setup
Control
Function
Setting Choices
Factory
Default
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...