Draft - 1 Aug 08
Positioning ECG Electrodes
6-16
ECG Monitoring
5-Wire
Placement
(EASI Mode)
Lead
Corresponds
to Standard
Lead
Placement
E
V
on the lower sternum at the level of the
fifth intercostal space
A
LL
on the left midaxillary line at the same
level as the E electrode
S
LA
on the upper sternum
I
RA
on the right midaxillary line at the
same level as the E electrode
N
Reference
can be anywhere, usually below the
sixth rib on the right hip
Note—
Make sure that the S and E electrodes line up vertically on
the sternum, and that the I, E and A electrodes align horizontally.
E
S
A
I