Configuring the System
Start-up and Configuration
8-13
M3170 Patient
Link
describes the test or inspection to perform for
each M3170 Patient Link
for each
type of test specified in
.
Note
Detailed procedures for performing the tests in the tables are given in the previous sections of
Testing Product Assurance.
Performance:
Verify purchased options are enabled.
Do appropriate data (waveforms and parameters) appear?
Verify normal system operation.
Do the devices respond to user inputs?
Do audible speaker tones occur?
Expected answers are
“Yes”.
If so,
Performance
test
passed.
P:P
or
P:F
where
P
= Pass
F
= Fail
If a Keyboard-Video-Mouse Switch is used, has the KVM
Switch successfully passed its
Performance
Test
?
Expected answers are
“Yes”.
If so,
Performance
test
passed.
PKVM:P
or
PKVM:F
where
P
= Pass
F
= Fail
If #C14 HL7 Export option is installed, verify successful
ping to the target HL7 client device by using hostname
Expected answer is
“Yes”. If so,
Performance
test passed.
PHL7:P
or
PHL7:F
where
P
= Pass
F
= Fail
If #C17 12-Lead Analysis/Export option is installed,
verify successful ping to the 12-Lead TraceMaster system
device by using hostname
Expected answer is
“Yes”. If so,
Performance
test passed.
P12LEAD:P
or
P12LEAD:F
where
P
= Pass
F
= Fail
If #C23 Holter Export option is installed, verify successful
Ping to Holter System by using hostname.
Expected answer is
“Yes”. If so,
Performance
test passed.
PHE:P
or
PHE:F
where
P
= Pass
F
= Fail
If #C65 Alert Data Export is installed, verify successful
ping to Alert Data Export device by using host name
Expected answer is
“Yes”. If so,
Performance
test passed.
PADE:P
or
PADE:F
where
P
= Pass
F
= Fail
Safety
Is the device connected to a 78581A/B (SCC) or telemetry
mainframe?
If answer is “No”, no
Safety
test required.
If answer is “Yes”,
move to next step.
S:NA
where
NA
= Not
required
Is there a 2nd earth ground connected to the SCC or
telemetry mainframe
Expected answer is
“Yes”.
If so,
Safety
test passed.
S:P
or
S:F
where
P
= Pass
F
= Fail
Confirm that 78581A/B or telemetry mainframe has
passed Safety Test Block (see
78581A/B
or
Telemetry
Mainframe
Test and Inspection Matrix
).
If answer is “Yes”,
Safety
test passed.
If answer is “No”,
move to next step.
S:P
or
S:F
where
P
= Pass
F
= Fail
Perform
78581A/B
or
Telemetry Mainframe Safety
Test Block
Expected answer is
“Yes”. If so,
Safety
test
passed
S:P
or
S:F
where
P
= Pass
F
= Fail
Table 8-3. M3150/M3155/M3151 - Test and Inspection Matrix
Test Block
Name
Test or Inspection to Perform
Expected Results
What to Record
on Service
Record
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...