Troubleshooting
Maintenance, Troubleshooting, and Repair
7-91
Third and Fourth
Character Codes
The third and fourth character identifies the:
•
Classification mode (single or multi-lead)
•
Detection mode (auto or manual)
•
User specified lead label (manual detection mode only)
•
User specified minimum threshold (150-350uV) (manual detection mode only)
•
Algorithm minimum detection threshold (150-350uV) (manual detection mode only)
Note
Use the Multi/Manual/Lead I entries as a guideline in determining the appropriate expanded
values for each Lead listed in the table below
Classification Mode
Detection Mode
User Specified
Detection Lead
User Specified
Detection Threshold
Algorithm Minimum
Detection
Threshold
Third and Fourth
Character
Multi
Auto
N/A
N/A
150
00
Manual
Lead I
150
150
350
10
11
a
200
200
350
12
13
a
250
250
350
14
15
a
300
300
350
16
17
a
350
350
350
18
19
a
Lead II
150 - 350
150 - 350
20 - 29
Lead III
150 - 350
150 - 350
30 - 39
Lead AVR
150 - 350
150 - 350
40 - 49
Lead AVL
150 - 350
150 - 350
50 - 59
Lead AVF
150 - 350
150 - 350
60 - 69
Lead V1
150 - 350
150 - 350
70 - 79
Lead V2
150 - 350
150 - 350
80 - 89
Lead V3
150 - 350
150 - 350
90 - 99
Lead V4
150 - 350
150 - 350
B0 - B9
Lead V5
150 - 350
150 - 350
C0 - C9
Lead V6
150 - 350
150 - 350
D0 - D9
Lead V
150 - 350
150 - 350
F0 - F9
Lead MCL1
150 - 350
150 - 350
G0 - G9
Summary of Contents for IntelliVue Series
Page 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Page 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Page 488: ...Procedure B 12 Web Installation on the Database Server ...
Page 492: ...Procedure C 4 External Modem Installation ...
Page 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...