background image

Philips IntelliVue Arm

InstallatIon InstructIons

P502673   Rev B

Summary of Contents for IntelliVue Arm

Page 1: ...Philips IntelliVue Arm Installation Instructions P502673 Rev B ...

Page 2: ... mounting holes positioned 6 center to center apart 8326 13 XXX 13 Wall Track 8326 19 XXX 19 Wall Track 8326 31 XXX 31 Wall Track NOTE Custom length tracks can be manufactured upon request 13 19 31 8328 XXX Track Mount The track mount slides into the wall track There is a set screw in the front of the mount which can be used as a locking mechanism and also controls the articulation of the mount ...

Page 3: ...um weight that can be placed on the arm is 32 lbs with a maximum weight capability of 53 lbs Setup includes 1 4 Allen wrench P701127 and 3 32 Allen wrench P700146 2 10 Adapter Assembly Options Select an adapter assembly size that is compatible with your application The Adapter Assembly Pole is available in two different lengths 61302 10 XXX 10 Pole Used for the IntelliVue adapter plug 61302 2 XXX ...

Page 4: ...IntelliVue Arm Assembly Instructions 4 ...

Page 5: ...stems resist a lateral force of 2 times the weight of the instrument and a vertical force of 2 2 times that weight applied simultaneously For 5 8 Gypsum Board walls to comply with CA seismic requirements for wall mounted instrumentation CA OSHPD Anchorage pre approval number TBA Anchorage or support of metal stud wall must be verified by the project engineer and approved by OSHPD prior to the inst...

Page 6: ...stems resist a lateral force of 2 times the weight of the instrument and a vertical force of 2 2 times that weight applied simultaneously For 5 8 Gypsum Board walls to comply with CA seismic requirements for wall mounted instrumentation CA OSHPD Anchorage pre approval number TBA Anchorage or support of metal stud wall must be verified by the project engineer and approved by OSHPD prior to the inst...

Page 7: ...stems resist a lateral force of 2 times the weight of the instrument and a vertical force of 2 2 times that weight applied simultaneously For 5 8 Gypsum Board walls to comply with CA seismic requirements for wall mounted instrumentation CA OSHPD Anchorage pre approval number TBA Anchorage or support of metal stud wall must be verified by the project engineer and approved by OSHPD prior to the inst...

Page 8: ...stems resist a lateral force of 2 times the weight of the instrument and a vertical force of 2 2 times that weight applied simultaneously For 5 8 Gypsum Board walls to comply with CA seismic requirements for wall mounted instrumentation CA OSHPD Anchorage pre approval number TBA Anchorage or support of metal stud wall must be verified by the project engineer and approved by OSHPD prior to the inst...

Page 9: ...se 3 16 Allen wrench P701311 and tighten the screws in the wall The length of the track you ve selected will depend on how many mounting holes there are and how many mounting screws P701274 you will need For this example a 19 track has been selected NOTE Do not fasten directly into sheetrock plaster or dry wall Consult the facilities personnel installer or your local hardware store if you have any...

Page 10: ...tach and secure mount to wall track Align the mount 8328 XXX in the grooves at the top of the wall track Slide it down to the desired height and tighten the set screws to lock into place using the 3 32 Allen wrench P700146 Track Mount Part No 8328 XXX ...

Page 11: ...IntelliVue Arm Assembly Instructions 11 Step 4 Insert monitor arm into track mount Insert the monitor arm 6958 XXX into the track mount 8238 XXX Track Mount Part No 8328 XXX Monitor Arm Part No 6958 XXX ...

Page 12: ...40 XXX up the pole and tighten the set screw to secure Slide the washer P700566 and retaining ring P700567 onto the pole Secure the retaining ring with retaining ring pliers Retaining ring pliers can be purchased at your local hardware store NOTE Attach the IntelliVue connector plug to the adapter and tighten using screws 1 4 20 x 1 2 Philips Flat Head Screw Part No P700987 Monitor Arm Part No 695...

Page 13: ...nagement P401136 Remove the backing and attach the first piece to the underside of the forearm This will keep your cords organized Be sure to leave enough slack for cables to loop around the elbow of the arm Remove the backing on the second piece and attach it to the underside of the lower channel Cable Managers Part No P401136 ...

Page 14: ...ll track cable cap P104985 and 1 wall mount cable cap P104599 Cut the track mount cable cap to desired length The wall track cable cap snaps into the wall track The track end caps simply plug into the top and bottom of the track Wall Track End Cap Part No P104596 Wall Mount Cable Cap Part No P104599 Wall Track Cable Cap Part No P104985 ...

Page 15: ...any non abrasive solution as approved by the facility in which the product will be used Stronger Weaker Stronger Weaker JOINT FRICTION Use the 3 32 Allen wrench P700146 to adjust 3 articulation points HEIGHT ADJUSTMENT Hold the arm in a horizontal position Use the 1 4 Allen wrench P700146 to strengthen or weaken Operating Manual Note The monitor must be attached to the arm to make any adjustments ...