Supported Modems
14
Chapter 2 - System Interfaces
The pinouts for the DB9 connector are shown in the following table.
The cable (M1380-61624) is a 3 meter long, fully wired RS-232 Null-modem
cable supplied with the viewer software. It is also available as an OB
TraceVue
option.
Supported Modems
Note—As modems are not designed to fulfil the electrical requirements of a
medical device in terms of leakage current and isolation, a modem must be
located outside of the patient vicinity.
The following modems are the only modems that are currently validated and
supported:
Table 1
Fully wired DB9 RS-232 Null-modem pinouts
Pin
Connects
to pin
Signal
Details
1 and 6
4
DTR
Data Terminal Ready
2
3
TxD
Transmit data to external devices
3
2
RxD
Receive data to external devices
4
1 and 6
DCD, DSR
Data Carrier Detect, Data Set
Ready
5
5
Gnd
RS-232 Ground
7
8
CTS
Clear To Send
8
7
RTS
Request To Send
9
9
-
Ring indicator, not connected