Repair Strategy
2
Chapter 1 - General Information
Who Should Read This Guide
This guide is for any technical personnel servicing and repairing the FM-2
monitor (M2922A).
You must:
• understand English
• be familiar with standard medical equipment installation procedures
• be familiar with current conventional technical terms as used throughout
this guide
What to Do Next
Familiarize yourself with the contents of this guide before attempting to service
or repair the monitor.
Repair Strategy
Reading operating error messages and the error log helps to determine whether a
fault is a hardware or software problem. Fault rectification is accomplished by
replacement of the defective part at the assembly level. For example, faults may
be repaired by replacement of a complete board where the fault is identified, or
by exchanging the monitor.
Repair or replacement of individual components on the boards is not supported,
and should never be attempted.
There are no servicing instructions in this manual for the Recorder (M2925A) as
a repair is effected by exchanging the unit only.
After any repairs you must carry out the following tests:
• Power-On Self-Test (see “FM-2 Power-On Self-Test” on page 20)
• Performance tests (see “Testing Transducers” on page 90)
• Safety tests after repair (see “Safety Testing” on page 84)
Summary of Contents for FM-2
Page 4: ......
Page 10: ......
Page 12: ......
Page 24: ...Routine Maintenance and Cleaning 12 Chapter 1 General Information ...
Page 42: ...Service Menu 30 Chapter 4 Power up Default Settings and the Service Menu ...
Page 74: ...Main PCB Disassembly Procedures 62 Chapter 6 Disassembly Reassembly Guide ...