Safety Testing
88
Chapter 8 - Test and Inspection Matrix
Safety Test 2: Patient Leakage Current Test
Normal
Condition
This test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied
Parts. The test measures patient leakage current from any individual patient
connection to earth (power ground) in parallel. It tests both normal and reversed
polarity.
Safety Test according IEC 601-1 (Clause 19).
Report the highest value.
Single Fault
Condition
(S.F.C.) Open
Supply
This test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied
Parts. The test measures patient leakage current from any individual patient
connection to earth (power ground) in parallel, with one supply lead
interrupted. It tests both normal and reversed polarity.
Safety Test according IEC 601-1 (Clause 19).
Report the highest value.
Test Steps
This test requires a sample patient cable for each device parameter. These must
be configured as recommended by the safety analyzer operating instructions.
1. Configure the electrical safety analyzer as recommended by the analyzer
operating instructions.
2. Connect the appropriate external power supply input power cord to the
analyzer as recommended by the analyzer operating instructions. Connect
the external power supply output cord to the monitor.
3. Turn the monitor on.
4. Perform the test as recommended by the analyzer operating instructions.
For each test condition, the measured leakage current must not exceed that
indicated in Table 23.
Summary of Contents for FM-2
Page 4: ......
Page 10: ......
Page 12: ......
Page 24: ...Routine Maintenance and Cleaning 12 Chapter 1 General Information ...
Page 42: ...Service Menu 30 Chapter 4 Power up Default Settings and the Service Menu ...
Page 74: ...Main PCB Disassembly Procedures 62 Chapter 6 Disassembly Reassembly Guide ...