6 Patient Alarms and INOPs
76
Technical Alarm Messages (INOPs)
INOP Message, Indication
Source
What to do
Batt EMPTY
INOP tone, battery LED flashes
during this INOP, alarms cannot
be paused or switched off.
Battery
The estimated remaining battery-powered operating time is less
than 10 minutes. Connect the monitor to AC immediately.
If the condition persists and the monitor is not connected to
mains power, this INOP is re-issued two minutes after you
acknowledge it.
Batt INCOMPAT.
INOP tone
Battery
The battery cannot be used with this monitor. Replace with the
correct battery.
Batt LOW
INOP tone
Battery
The estimated battery-powered operating time remaining is less
than 20 minutes.
Batt MALFUNCTION
INOP tone, battery LED flashes
during this INOP, alarms cannot
be paused or switched off unless
the monitor is connected to
mains power.
Battery
The monitor cannot determine the battery status. If this INOP
persists, replace the faulty battery. If the condition persists and
the monitor is not connected to mains power, this INOP is
reissued two minutes after you acknowledge it.
Batt MISSING
Battery
No battery found in FM20/30. Contact your service personnel.
BUS MASTER MALFUNC
Monitor
There is a problem with the monitor's hardware. Contact your
service personnel.
CHARGER MALFUNC.
INOP tone, battery LED may
flash
Battery
There is a problem with the battery charger in the monitor.
Connect the monitor to mains power and contact your service
personnel.
CHECK BATT TEMP
Battery
The temperature of the battery is too high. Check that the
monitor is not exposed to heat.
Check Flex Texts
INOP tone
Monitor
If this INOP appears, check the monitor and patient settings
before you resume monitoring. If the settings are unexpected,
there may be a problem with the monitor software. Contact your
service personnel.
Check Monitor Func
INOP tone
Monitor
A potential internal problem with the monitor has been
detected. Contact your service personnel.
Check Keyboard
INOP tone
Monitor
Perform a visual and functional check of the keyboard. Contact
your service personnel.
Check Mouse Device
INOP tone
Monitor
Perform a visual and functional check of the mouse input
device. Contact your service personnel.
CHECK PAPER
INOP tone
Recorder
Check that there is no paper jam, that the print drawer is
properly shut, that the paper is loaded with the grid facing
upwards, and that the correct Philips paper is being used.
Check Settings
INOP tone
Monitor
If this INOP appears, check the monitor and patient settings
before you resume monitoring. If the settings are unexpected,
there may be a problem with the monitor software. Contact your
service personnel.
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...