5 Alarms
71
Testing Alarms
In general, to test the functioning of visible and audible alarms, do the following:
1
Enable the alarm.
2
Set the alarm limits.
3
Measure or simulate the parameter that is out of range, or signal loss.
4
Verify that the visible and audible alarms are working.
As an example, to test the FHR alarms:
1
Connect the US transducer to a fetal sensor socket.
2
Enable the FHR alerting (see “Turning Alarms On or Off ” on page 127).
3
Set the high alert limit and delay to 150 bpm and 60 seconds respectively, and the low alert limit
and delay to 110 bpm and 60 seconds respectively (see “Changing Alarm Limits” on page 127).
4
Generate a fetal heart rate of approximately 180 bpm (3 beats per second) for more that one
minute.
5
Verify the functioning of the visible and audible alarm.
Alarm Behavior at Power On
Selecting
AlarmsOffAtStart
will cause alarms to be initially suspended or off the next time the monitor
is switched on (depending on the setting for Alarms Off). The Alarms Off as Infinite behaves in the
same manner as when Alarm Mode equals INOP Only.
In order for alarms to be suspended or switched off initially, the monitor must be switched off for
more than one minute, the last main alarm state was set to off or suspended.
The exception to this condition is when the alarms off state has been set to Infinite, and the new active
setting is set to Infinite.
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...