26 Specifications and Standards Compliance
208
Electromagnetic Emissions and Immunity
The EMC standards state that manufacturers of patient-coupled equipment must specify immunity
levels for their systems. See Tables 1 to 4 for this detailed immunity information. See Table 5 for
recommended minimum separation distances between portable and mobile communications
equipment and the product.
Immunity is defined in the standard as the ability of a system to perform without degradation in the
presence of an electromagnetic disturbance.
Caution should be exercised in comparing immunity levels between different devices. The criteria used
for degradation are not always specified by the standard and can therefore vary with the manufacturer.
In the table below, the term "device" refers to the Avalon FM20/30/40/50 fetal monitor together with
its accessories. The table gives details of the electromagnetic emissions, and how these are classified,
for the device, and the electromagnetic environments in which the device is specified to technically
function.
1
Note that the device is not intended for home use.
Table 1 - Guidance and Manufacturer's Declaration: Electromagnetic Emissions
Emissions Test
Compliance
Avoiding Electromagnetic Interference
Radio Frequency (RF) emissions
Group 1
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations and flicker IEC 61000-3-3 complies
RF emissions CISPR 11
For the Avalon FM20/30 fetal monitor with all
accessories except the IUP/ECG patient
module M2738A.
Class B
The device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage supply network that
supplies buildings used for domestic purposes
1
.
RF emissions CISPR 11
For the Avalon FM40/FM50 with all
accessories.
For the Avalon FM20/30 fetal monitor
whenever used with the IUP/ECG patient
module M2738A.
For the Avalon CTS Interface Cable (M2731-
60001/M2732-60001) whenever used with the
Avalon CTS Cordless Fetal Transducer System.
Class A
The device is suitable for use in all
establishments other than domestic
establishments and those directly connected to
the public low-voltage supply network that
supplies buildings used for domestic purposes.
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...