19 Monitoring Maternal Heart / Pulse Rate
160
Adjusting the Heart Rate / Pulse Alarm Limits
To adjust the pulse alarm limits:
1
In the
Setup SpO
₂
menu, select
Pulse (SpO
₂
)
. This opens the
Setup Pulse (SpO
₂
)
menu.
2
Ensure
Pulse (SpO
₂
)
is
On
. Select
Pulse (SpO
₂
)
to toggle between
On
and
Off
.
3
Set the pulse alarm limit:
– Select
High Limit
then choose the upper alarm limit for tachycardia from the pop-up list.
– Select
Low Limit
then choose the lower alarm limit for bradycardia from the pop-up list.
Average Pulse Rate from Noninvasive Blood
Pressure
WARNING
No alarm is possible when noninvasive blood pressure is the source of the pulse rate.
When you are measuring noninvasive blood pressure, the monitor can also calculate the average pulse
rate. This occurs in either manual or automatic mode, when neither MECG, SpO
2
nor pulse from
Toco MP is being measured. The value is displayed on the screen, and printed on the trace. It is not the
actual pulse value, but an average pulse rate, taken during the most recent noninvasive blood pressure
measurement. The value is updated after each successive measurement. If you need a continuous
measurement, you should monitor using MECG, SpO
2
or pulse from Toco MP.
Troubleshooting
Testing MECG Mode
Refer to the monitor's
Service Guide
.
Problem
Possible Causes
Solutions
MECG LEADS OFF
displayed
.
Numeric is displayed with
a - ? - for 10 seconds;
INOP tone.
See also “Patient Alarms
and INOPs” on page 73.
One or more MECG leads is not
attached.
Make sure that all required leads are attached.
Bad electrical contact.
Check positioning of the electrode, ensuring that
none are displaced.
Check electrodes and replace if necessary.
Electrodes defective.
prints repeatedly
The ultrasound transducer is
measuring maternal pulse.
Reposition the ultrasound transducer.
MECG EQUIP MALF
displayed.
See “Patient Alarms and INOPs” on page 73.
MECG UNPLUGGED
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...