17 Monitoring Noninvasive Blood Pressure
147
NOTE
Be aware that a combination of a recorder speed of less than 3 cm/min and a repetition time of less
than five minutes can result in not all noninvasive blood pressure measurements being recorded on the
fetal trace. For example, if the recorder speed is set to 1 cm/min and the repetition time is set to
2 minutes, due to the low speed setting, the recorder will only be able to record every other
noninvasive blood pressure measurement. This affects only the local fetal trace recording, and all
measurements are displayed as normal on the monitor's screen.
Choosing the Alarm Source
You can monitor for alarm conditions in systolic, diastolic and mean pressure, either singly or in
parallel. Only one alarm is given, with the priority of mean, systolic, diastolic.
If mean is not selected as alarm source (
Sys.
,
Dia.
, or
Sys & Dia
selected), but the fetal monitor can
only derive a mean value, mean alarms will nevertheless be announced using the most recent mean
alarm limits. Check that the mean alarm limits are appropriate for the patient, even when not using
mean as the alarm source. When no value can be derived an
NBP MEASURE FAILED
INOP will be
displayed.
Assisting Venous Puncture
You can use the cuff to cause sub-diastolic pressure. The cuff deflates automatically after a set time if
you do not deflate it.
1
In the
Setup NBP
menu select
VeniPuncture
.
2
Puncture vein and draw blood sample.
3
Reselect
VeniPuncture
to deflate the cuff.
During measurement, the display shows the inflation pressure of the cuff and the remaining time in
venous puncture mode.
Menu option
Pressure value monitored
Sys.
systolic
Dia.
diastolic
Mean
mean
Sys & Dia
systolic and diastolic in parallel
Dia & Mean
diastolic and mean in parallel
Sys & Mean
systolic and mean in parallel
Sys&Dia&Mean
all three pressures in parallel
1
Cuff pressure
2
Venous puncture measurement
mode
3
Time left in venous puncture
mode
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...