Testing the Parameter Signals
Chapter 3 - Tests and Error Messages
Testing the Parameter Signals
The parameter test checks the signal path to and from the transducer sockets,
but not the transducers themselves.
1. Switch on the monitor, the recorder and the telemetry receiver.
2. Connect the appropriate transducer to each socket on the transmitter.
3. Standing within view of the monitor, press the Nurse Call Button on
the transmitter and switch on the transmitter simultaneously. The test
runs while the Nurse Call is pressed.
If the response is different, see troubleshooting flowcharts, Chapter 4.
Signal
Correct Monitor Response
US
125
is displayed and printed.
Signal Quality Indicator is green.
Fetal heartbeat is heard from the loudspeaker.
Toco
A triangle signal with an amplitude of
40
units is
displayed and printed.
Each cycle lasts for 12 seconds.
DECG
150
is displayed and printed
Signal Quality Indicator is green.
Fetal heartbeat is heard from the loudspeaker.
Summary of Contents for 50 T Series
Page 4: ......
Page 10: ...x List of Figures ...
Page 26: ...Error Messages Chapter 3 Tests and Error Messages ...
Page 36: ...TOCO Channel Chapter 4 Troubleshooting TOCO Channel Figure 8 Troubleshooting TOCO Channel ...
Page 39: ...Event Marker Chapter 4 Troubleshooting Event Marker Figure 11 Troubleshooting Event Marker ...
Page 40: ...Nurse Call Chapter 4 Troubleshooting Nurse Call Figure 12 Troubleshooting Nurse Call ...
Page 42: ... Chapter 5 Interfacing to a Fetal Monitor ...
Page 48: ...Using the Service Program Chapter 6 Using the Service Software ...
Page 51: ...Lists of Parts Chapter 7 Replacing Parts Figure 14 Transmitter Parts ...
Page 53: ...Lists of Parts Chapter 7 Replacing Parts Figure 15 Receiver Boards ...
Page 78: ...Receiver VCXO Test Chapter 8 RF Bench Repair ...