1
MRM10E
User’s Guide
Digital Micro-Ohmmeter
GU-1522 Copyright © Phenix Technologies. All rights reserved.
Rev 6/22/2018
Page 1: ...1 MRM10E User s Guide MRM10E Digital Micro Ohmmeter GU 1522 Copyright Phenix Technologies All rights reserved Rev 6 22 2018...
Page 2: ...2 MRM10E WARNING Complete Grounding of this unit is necessary for the safe operation of this equipment Disconnect inputs before ungrounding this equipment...
Page 3: ...t the Control Panel to have access to the internal parts is dangerous as there are high voltages inside capable of causing fatal accidents Cleaning of this instrument should be carried out using a sof...
Page 4: ...Operating principle 5 2 Control panel 6 3 Power supply 7 3 1 Battery condition 7 3 2 Battery charger 7 4 Measurement 8 5 Messages 9 6 Some notes about accuracy 9 7 RS232 output 10 8 Cleaning 10 9 Repl...
Page 5: ...od Digital reading alphanumeric display Up to 4 digits readings Powered by rechargeable battery or mains supply 0 1 resolution 200 maximum reading 1 1 Operating Principle This device uses the Kelvin B...
Page 6: ...5 STOP Key 4 Ground 16 START Key 5 Current Terminal C 17 Hold Key retains last measurement in display 6 Potential Terminals P P 7 Current Terminal C 8 ON OFF Switch 9 Displays both measured resistance...
Page 7: ...chieves a critical level the display will show the BAT message notifying that the charge level is low After a few minutes the measurement will be automatically interrupted to preserve the battery from...
Page 8: ...ncreases the current Adjust this control until obtaining the desired value or until the current indicator bar graph indicates 100 The bar graph will show the test current value as a percentage of the...
Page 9: ...igher than the maximum value readable in the selected range H Indicates that the value is the one retained in the memory when pressing the hold key BAT It shows that the battery is discharged It is ne...
Page 10: ...uR micro ohm mR milli ohm R ohm 8 Cleaning CAUTION Before cleaning ensure that the unit is Off and removed from power connection The panel terminals and connectors of the equipment must stay dry and...
Page 11: ...al data output RS 232 4800 bps Suitable for data collection in an external serial printer computer or data logger Environmental protection IP54 with closed lid Safety class Meets the requirements of I...
Page 12: ...or s Manual are correct If there are concerns or comments as you have used this information Phenix Technologies appreciates any feedback Unit Serial Number SECT PAGE S COMMENT Please return to Phenix...