
UNITS:
Selectable: mm or inch.
UPDATE
RATE:
Update the rate of measurement result. Optional 4, 8 or 16Hz.
AUTO
POWER
DOWN:
The device can be set to shut off after 5, 10, 20 minutes of inactivity. If set to OFF then the
Cal/ON
button will power the unit on or off when pressed.
DELETE
ALL
FILES:
Empty all memory of values and waveforms.
DEFAULT
SETUP
: Brings unit back to original factory settings
3.3
CALIBRATION
Before using the UTG‐4000, calibrating the gage as well as the probe is necessary. See the following
steps:
3.3 CALIBRATION STEP ILLUSTRATION
1
. Adjust the sound velocity to 0.233” or (5900 metric)according to the method in chapter 3.1.
2
. After turning on the gage, press to enter the calibration interface. The screen shows
PUT
PROBE
ON
ZERO
BLOCK
USE
COUPLANT
.
3
. Spread the couplant according to the presentation, then put the probe on the zero block and make it
coupled completely until the screen shows
ACQUIRING
PLEASE
STAY
COUPLED.
4
. Hold a few seconds, and the screen will display
PROBE
ZERO
COMPLETE
REMOVE
PROBE
FROM
ZERO
BLOCK.
Remove the probe from the zero block and the display returns back to the measuring
interface automatically.
The
display
should
show
0.157”
or
4mm.
If
any
other
values
appear
then
you
must
adjust
the
sound
velocity
accordingly.
5
. During the calibration process, press the “P2” button to ABORT at any time to stop the calibration
process and the unit will go back to the measuring interface.
Summary of Contents for UTG-400
Page 22: ......