4.
DATA
STORAGE
FUNCTIONS
The UTG‐4000 utilizes a micro grid format allowing this unit to store up to 100,000 thickness values
and 1,000 A/B scan waveforms.
See
Fig
3.17
below
Thickness values and waveforms can be mixed and stored in the same file. The stored data can be
transferred to your PC via USB output and saved as an EXCEL or TXT file.
4.1
THICKNESS
VALUE
AND
WAVEFORM
STORAGE
1)
Storage file number
2)
Line mark
3)
Row mark
4)
Data in memory
5)
Return to main menu
6)
Save current value or waveform
7)
Delete the selected data
4.2
Browsing
stored
data
Viewing
Individual
Files:
Press “M” to enter the configuration display. Highlight “GRID FILE” and press the P1 button to view the
file numbers. Use the up and down arrows to choose the file number you want to recall. Press the P2
button to confirm and view the individual file.
Viewing
All
Files
in
a
specific
group:
Press “M” to enter the configuration display. Highlight “GRID FILE” and press the P2 button to view the
all files within that group number that is highlighted. Use the up and down arrows to choose the file
number within that group that you want to recall.