measured, the gauge is keeping track of the lowest measurement it finds. The transducer may be “scrubbed”
across a surface, and any brief interruptions in the signal will be ignored. When the transducer loses contact with
the surface for more than two seconds, the gauge will display the smallest measurement it found. When the
transducer is removed from the material being scanned, the gauge will display the smallest measurement it
found.
When the scan mode is turned off, the single point mode will be automatically turned on. Turn on/off the
scan mode by the following steps:
Press the
key to switch the scan measurement mode on and off. It will display the current condition of
the scan mode on the main screen.
4.6 Changing Resolution
UTG-2800 has selectable display resolution, which is 0.01” / 0.001” (0.1mm and 0.01mm).
Press down the
key while turning on the gauge will switch the resolution between “High” and “Low”.
4.7 Changing Measuring Units
On the measurement mode, press the
key to switch back and forth between imperial and metric units.
4.8 Memory Management
4.8.1 Storing a reading
There are twenty files (F00-F19) that can be used to store the measurement values inside the gauge. At
most 100 records (thickness values) can be stored to each file. By simply pressing the
key after a new
measurement reading appears, the measured thickness value will be saved to current file. It is added as the last
record of the file. To change the destination file to store the measured values, follow the steps:
1) Press the key to activate the data logging functions. It will display the current file name and the total
record count of the file.
2) Use
the
key and the
key to select the desired file to set as current file.
3) Press
the
key to exit the data logging functions at any time.
4.8.2 Clearing selected file
The user may require the contents of an entire file be completely cleared of all measurements. This would
allow the user to start a new list of measurements starting at storage location L00. The procedure is outlined in
the following steps.
1. Press
the key to activate the data logging functions. It will display the current file name and the total
record count of the file.
2. Use
the
key and the
key to scroll to the file that will be cleared of all measurements.
3. Press
the key on the desired file. It will automatically clear the file, and display “-DEL”.
4. Press
the key, at any time, to exit the data logging functions and return to measurement mode.