Troubleshooting/Options
Troubleshooting
Optional Accessories: SRG-4500
1) Precision Support Stand
2) Stylus Extension
3) Special stylus for round and grooved surfaces
Error messages
Possible reasons
Troubleshooting
Out of range!
Maximum value of measured
signal exceeds measuring range;
1
.
Press Esc key to return;
2
.
Enter menu setting status,
increase measuring range,
press Esc key to return;
3
.
Measure again.
No data!
Wrong operation results in failed
measuring;
1
.
Press Esc key to return;
2
.
Check if preparation for
measuring is correct;
3.Switch on and measure again.
A/D failure
Hardware circuit fault;
1: Switch off and switch on
again;
2: Press Reset key;
3: Return to Phase II for repair.
Motor failure
Mechanical fault;
1: Switch off and switch on
again;
2: Press Reset key;
3: Return to Phase II for repair.
Pickup failure
Pickup is in automatic return
process
Press Esc key to return and wait
till pickup returns to the start
position