) MAJOR TECHNICAL PARAMETERS
ROUGHNESS
PARAMETERS
Ra, Rz, Rq, Rt
MEASURING
RANGE
RA:
0.05
‐
10.0µM
RZ:
0.1
‐
50µM
Rq :
0.05
~
10.0µm
Rt:
0.1
~
50µm
CUT
‐
OFF
LENGTHS
0.25, 0.80, 2.50
FILTER
RC ANALOG
TRACING
LENGTH
0.23IN (6MM)
TRACING
SPEED
0.04IN/SECOND (1.0MM/SECOND)
ACCURACY
+/‐12% of known value
PICK
UP
STYLUS
PIEZO‐ELECTRIC
TRACER
TIP
DIAMOND, RADIUS 10µM +/‐ 1µM
OPERATING
TEMPERATURE
32‐104 DEGREES F (0‐40 DEGREES C)
POWER
3.7v Lithium ion Battery
CONTACT
FORCE
ON
PROBE
<0.5N
STATIC
MEASURING
FORCE
OF
SENSOR
STYLUS
< 0.016N
DIMENSIONS
106 X 70 X 24mm
WEIGHT
0.4LBS (200G)
4)
USE
AND
OPERATION
4.1)
OPERATION:
POWER ON THE DEVICE.
Press and hold the Red Power button on the front panel for 1 second. After a “beep” sound, the device
is ready to work. The large LCD display will show the measuring parameters and sampling length of the
previous test. Before taking a test, choose the desired parameter
Ra
,
Rz
,
Rq
, or
Rt
and proper sampling length
2.5
,
0.8
or
0.25
(for sampling length option, consult the appendix).
After switching on the device
,
press the to change parameters.
To choose the proper sampling length, press the keypad
(
0.25
,
0.8
OR
2.5
).
After the parameters and sampling lengths are taken, you may begin testing. Position the
l l
mark over the test area and press the start button on top of the unit. The stylus moves
automatically. When the device beeps twice, the measurement has been completed and the results will be
shown upon the LCD screen.