22
Table 3
Type of impact device
DC(D)/DL
D+15
C
G E
Impacting energy
Mass of impact body
11mJ
5.5g/7.2g
11mJ
7.8g
2.7mJ
3.0g
90mJ
20.0g
11mJ
5.5g
Test tip hardness:
Dia. Test tip:
Material of test tip:
1600HV
3mm
Tungsten
carbide
1600HV
3mm
Tungsten
carbide
1600HV
3mm
Tungsten
carbide
1600HV
5mm
Tungsten
carbide
5000HV
3mm
synthetic
diamond
Impact device diameter:
Impact device length:
Impact device weight:
20mm
86(147)/
75mm
50g
20mm
162mm
80g
20mm
141mm
75g
30mm
254mm
250g
20mm
155mm
80g
Max. hardness of
sample
940HV 940HV 1000HV 650HB 1200HV
Mean roughness value
of sample surface Ra:
1.6
μ
m 1.6
μ
m 0.4
μ
m 6.3
μ
m 1.6
μ
m
Min. weight of sample:
Measure directly
Need support firmly
Need coupling tightly
>5kg
2
~
5kg
0.05
~
2kg
>5kg
2
~
5kg
0.05
~
2kg
>1.5kg
0.5
~
1.5kg
0.02
~
0.5kg
>15kg
5
~
15kg
0.5
~
5kg
>5kg
2
~
5kg
0.05
~
2kg
Min. thickness of sample
Coupling tightly
Min. layer thickness for
surface hardening
5mm
≥
0.8mm
5mm
≥
0.8mm
1mm
≥
0.2mm
10mm
≥
1.2mm
5mm
≥
0.8mm
Size of tip indentation
Hardness
300HV
Indentation
diameter
Depth of
indentation
0.54mm
24
μ
m
0.54mm
24
μ
m
0.38mm
12
μ
m
1.03mm
53
μ
m
0.54mm
24
μ
m
Hardness
600HV
Indentation
diameter
Depth of
indentation
0.54mm
17
μ
m
0.54mm
17
μ
m
0.32mm
8
μ
m
0.90mm
41
μ
m
0.54mm
17
μ
m
Hardness
800HV
Indentation
diameter
Depth of
indentation
0.35mm
10
μ
m
0.35mm
10
μ
m
0.35mm
7
μ
m
--
--
0.35mm
10
μ
m
Available type of impact
device
DC: Test hole or
hollow
cylindrical;
DL:Test narrow
groove or hole
D
+15:
Test
grooves
C: Test
small,light,thi
n parts and
surface of
hardened
layer
G:
Test
large, ,hea
vy and
rough
surface
steel
E:
Test super
high
hardness
material