Display Lighting
Display lighting is switched on and off by pushing
Ï
button from the first level in the “Scale” menu.
Operation with display lighting will slightly reduce the time the device can operate on a full charge.
Operation with Electronic Module
Choosing parameters
Power Unit on according to instructions:
Scale Selection
•
Begin by pressing the
Ï
button to get to the – ‘scale’ – parameter. In the display upper part,
words ‘XXX scale’ will appear, where ‘XXX’ can be HRC, HRB, HB, etc
•
Select the hardness scale you need by pushing
Í
and
Î
buttons.
•
Confirm the selection of the required for you hardness scale by pushing
Ð
button. After that
you will automatically advance to the second level “TEST”
Operation Mode Selection
•
Select the required operation mode – ‘TEST’ or ‘CALIBRATE’, or ‘PROCESS’
by pushing either
Í
and
Î
buttons.
•
when you are in the parameter that you need, press the
Ð
button to confirm this procedure.
The hardness tester is now ready for that operation.
Operation Canceling
To cancel an operation or to return to the previous one press the
Ï
button.
‘Test’ Mode
‘Test’ mode and all operations within this mode shall be conducted separately for U1 ultrasonic
transducer and D1 dynamic transducer. The given below operations can be performed in this mode.
•
Archive
•
Measurement and recording
‘Archive’ Operation
Hardness tester operation in ‘measurement’ mode always starts with the ‘Archive’ operation.
Attention!
During the first operation it is recommended to skip the ‘archive’ operation and proceed with
‘measurement and recording’ operation. For this purpose, push
Ð
button.