Maintenance and Repair
Avoid placing the MET near locations where it can be exposed to shock, heavy dust, dampness, strong
magnetic fields and any form of lubricious fluids.
Battery Replacement:
•
Remove two rubber plugs on the back side Bottom half of the MET-U1A. Use the supplied hex
wrench to loosen the two hex screws. Lift the metal battery cover off to expose the 4-AA
batteries. It is recommended that you replace the batteries with a high quality NiMh set of
rechargeable batteries. Standard AA batteries can also be used but life expectancy will not be
very long. Pay careful attention to the polarity when installing new batteries. Please be sure and
dispose of the NiMh batteries properly and according to local recycling laws.
Impact Device D
•
Using the included brush, clean the guide tube and impact body after it has performed 1000-2000
tests using the following procedure:
Screw off the support ring and take out the impact body.
Clean the tube by winding the brush counter-clockwise into the
tube until it reaches the bottom and then pull it out. Repeat this
procedure 5-6 times.
Replace the impact body and support ring.
Release the impact spring after each test.
Never apply lubrication of any kind to
impact body or tube!
Repair
Dynamic impact device
•
When the tester is calibrated against the standard test block and results are out of tolerance, it
could
mean the carbide ball inside of the impact device is worn out and should be replaced.
•
Upon coming across any functional problems not mentioned in this manual, contact a Phase II
Customer Service Representative at (201)933-6300.
•
DO NOT
open the
MET
or dismount any of its fixed parts. This precision instrument should
only be serviced by a factory trained service representative.
•
If the unit is opened, its warranty will automatically become void
.