phase II+ 900-355 Manual Download Page 9

 

7

5.3 Testing 

Caution

The sample must be place and supported in such a manner that the surface to 

be indented is in a plane perpendicular to the axis of the indenter and the line of the 
indenting force. Otherwise 3000kgf will bring a great divisive force in horizontal direction 
that may cause damage to the machine. 

 
 
 
 
 
 

      YES                            NO 

 

 

 

 

Place the sample to be tested on the anvil, and rotate the hand wheel in clockwise to lift the anvil 
as figure 5.3.1, showing the anvil moving course. Rotate the hand wheel smoothly until the anvil 
in figure reaches the end position, buzzing warning will be given out, then the rotating should be 
stopped at right. 

Figure 5.3.1                                                            Figure 5.3.2 

 
 

Then the testing cycle will be performed automatically as following:    load application showing as 
figure 5.3.2, the load holding showing as figure 5.3.3 with dwell time counting down, finally 
unloading is performed immediately when the dwell time has finished.   
After that, all the test results will be shown on the display as in figure 5.3.4.   
Rotate the hand wheel in counter-clockwise to lower the anvil, and the screen returns to figure 
5.1.1 for the next test.   
 
 
 

Summary of Contents for 900-355

Page 1: ...Saddle River NJ 07458 Tel 201 962 7373 Fax 201 962 8353 E Mail info phase2plus com Web Site http www phase2plus com Model No 900 355 Model No 900 356 Shown w Optional Optical Scanner System PHT 5000...

Page 2: ...ructure schematic diagram 3 4 Installation and adjustment 3 5 Operating methods 4 5 1 Testing preparation 4 5 2 Test parameters setting 5 5 2 1 Brinell hardness symbol 5 5 2 2 Load dwell setting 6 5 2...

Page 3: ...29400N 1500Kgf 14700N 1000Kgf 9800N 750Kgf 7355N 500Kgf 4900N 250Kgf 2452N 187 5Kgf 1839N 125Kgf 1226N 100Kgf 980N 62 5Kgf 612 9N Load dwell duration 2s 99s can be set and stored Tungsten Carbide Ball...

Page 4: ...Ball indenter 1 5mm Tungsten Carbide Ball indenter 1 2 5mm Tungsten Carbide Ball indenter 1 Mounting screws for indenter 1 Screwdriver for indenter mounting 1 Flat anvil 1 V shape anvil 1 20X microsco...

Page 5: ...provide working clearance for the leadscrew refering figure 4 1 4 3 Place the tester on the prepared bench turn the hand wheel counter clockwise to lower the anvil remove the anti vibration pad then...

Page 6: ...rinell testing parameter showing the testing load indenter diameter load dwell duration as well as current time Figure 5 1 1 The proper indenter must be mounted in the machine according to the specifi...

Page 7: ...ars press to confirm At this time test force will be changed automatically Press Setup key to return to figure 5 1 1 or press or key to reset the other parameters Tab 5 1 Brinell hardness parameters H...

Page 8: ...ress key then the figure 5 2 3 and 5 2 4 appear which showing the directions for 0 102F D2 ratios should be selected according to the materials and the range of hardness value Figure 5 2 3 Figure 5 2...

Page 9: ...moving course Rotate the hand wheel smoothly until the anvil in figure reaches the end position buzzing warning will be given out then the rotating should be stopped at right Figure 5 3 1 Figure 5 3 2...

Page 10: ...lens Adjusting Knob Reading Taking Drum 1 2 Application This microscope is designed to be used primarily for measurement of the indent formed in Brinell hardness test 1 3 Technical Specification Ampli...

Page 11: ...um with the scale span of 0 005 mm so that a complete turn of the drum equals to 0 5 mm See Fig 9 for details 1 4 3 There are two graduation scales in the eye lens one fixed and one mobile Numerals fr...

Page 12: ...e restored 6 2 When performing any adjustments and examining in the machine the power supply must be turned off 6 3 When installing or removing the indenter the mounting surface should be clean and dr...

Page 13: ...atic measurement Field proven under severe sample surface conditions 3 Single camera with four magnifications 1 3M pixel CMOS USB camera with two camera tubes with each tube height having two magnific...

Page 14: ...atic measurement Automatically or manually measures the indentation diameters on the calibration sample and the test sample and automatically interpolates the HB value for the test sample 4 Conversion...

Page 15: ...18 311 304 297 290 284 278 85 6 85 0 84 5 83 9 83 4 82 8 82 3 81 8 81 2 80 7 80 1 79 6 79 0 78 5 78 0 77 4 76 8 76 3 75 9 75 2 74 7 74 1 73 6 73 1 72 5 72 0 71 5 70 9 70 4 69 9 69 4 68 9 68 4 67 9 67...

Page 16: ...236 231 226 221 216 211 206 201 196 192 188 184 180 176 173 170 167 164 161 158 155 152 150 147 145 143 141 139 137 135 131 129 127 125 124 61 5 60 9 60 2 59 5 58 9 58 3 57 6 57 0 56 4 55 8 55 2 54 6...

Page 17: ...611 603 596 590 583 576 570 563 557 551 545 539 533 527 521 516 510 62 61 60 59 58 57 56 55 54 53 52 51 50 49 48 47 46 45 44 43 42 41 40 39 38 37 36 35 34 33 32 31 30 29 28 27 26 25 24 23 22 21 20 76...

Page 18: ...s Azcapotzalco C P 02020 Mexico Tel 01 55 3622 7000 Fax 01 55 5319 4000 BEIJING CHINA Phase II Measuring Instruments Beijing Ltd Room 301 Bldg 2 Qing Yuan Xi Li Haidian District Beijing 100192 China T...

Reviews: