phase II+ 900-355 Manual Download Page 12

 

10

Measure the diameter of the indention using the microscope in 2 directions perpendicular to each 
other. Remember the values and enter as shown below, marked as d1 and d2 on the digital 
display of the machine 
Press “ ”or“ ” key to move cursor to “3” HBW Calculator in figure 5.2.1, then press “

” key 

to confirm. Figure5.3.5 appears. Press “ ”or“ ”key to input the two diameters, d1 and d2. And 
press “

” key to confirm. The Brinell hardness value will be calculated automatically.   

Figure 5.3.5 

5.4 Shut down 

Remove the test force completely, and switch off the power supply. Remove power cord when machine will not 

be in use for an extended period of time. 

6 Maintenance   

6.1

 When the hardness tester is to be repositioned in another area of your facility, the indenter must be 

removed. If it is to be transported for a long distance, the original packaging must be restored. 

6.2

 

When performing any adjustments and examining in the machine, the power supply must be turned 

off. 

6.3

 When installing or removing the indenter, the mounting surface should be clean and dry. The indenter 

should be removed carefully and placed back in its supplied case if not being used for an extended period of 
time. 

6.4

 The surface of anvil and standard hardness blocks should be clean without contamination, scratches or 

rust. When storing anvils or test blocks, they should be lightly coated with oil to prevent rusting.   

6.5

 Dustproof and corrosive medium prevention should be considered in daily operating environment. Rust 

prevention measures should be adopted in high humidity area. 

6.6

 The leadscrew of the anvil should be lubricated periodically. Apply a few drops of light machine oil, 

then run the leadscrew up and down a few times to distribute the oil.   

6.7

 Please don’t disassemble or adjust any fixed parts. This will void your warranty. This should be done 

by a professional service person familiar with the machine.   

 

 

Summary of Contents for 900-355

Page 1: ...Saddle River NJ 07458 Tel 201 962 7373 Fax 201 962 8353 E Mail info phase2plus com Web Site http www phase2plus com Model No 900 355 Model No 900 356 Shown w Optional Optical Scanner System PHT 5000...

Page 2: ...ructure schematic diagram 3 4 Installation and adjustment 3 5 Operating methods 4 5 1 Testing preparation 4 5 2 Test parameters setting 5 5 2 1 Brinell hardness symbol 5 5 2 2 Load dwell setting 6 5 2...

Page 3: ...29400N 1500Kgf 14700N 1000Kgf 9800N 750Kgf 7355N 500Kgf 4900N 250Kgf 2452N 187 5Kgf 1839N 125Kgf 1226N 100Kgf 980N 62 5Kgf 612 9N Load dwell duration 2s 99s can be set and stored Tungsten Carbide Ball...

Page 4: ...Ball indenter 1 5mm Tungsten Carbide Ball indenter 1 2 5mm Tungsten Carbide Ball indenter 1 Mounting screws for indenter 1 Screwdriver for indenter mounting 1 Flat anvil 1 V shape anvil 1 20X microsco...

Page 5: ...provide working clearance for the leadscrew refering figure 4 1 4 3 Place the tester on the prepared bench turn the hand wheel counter clockwise to lower the anvil remove the anti vibration pad then...

Page 6: ...rinell testing parameter showing the testing load indenter diameter load dwell duration as well as current time Figure 5 1 1 The proper indenter must be mounted in the machine according to the specifi...

Page 7: ...ars press to confirm At this time test force will be changed automatically Press Setup key to return to figure 5 1 1 or press or key to reset the other parameters Tab 5 1 Brinell hardness parameters H...

Page 8: ...ress key then the figure 5 2 3 and 5 2 4 appear which showing the directions for 0 102F D2 ratios should be selected according to the materials and the range of hardness value Figure 5 2 3 Figure 5 2...

Page 9: ...moving course Rotate the hand wheel smoothly until the anvil in figure reaches the end position buzzing warning will be given out then the rotating should be stopped at right Figure 5 3 1 Figure 5 3 2...

Page 10: ...lens Adjusting Knob Reading Taking Drum 1 2 Application This microscope is designed to be used primarily for measurement of the indent formed in Brinell hardness test 1 3 Technical Specification Ampli...

Page 11: ...um with the scale span of 0 005 mm so that a complete turn of the drum equals to 0 5 mm See Fig 9 for details 1 4 3 There are two graduation scales in the eye lens one fixed and one mobile Numerals fr...

Page 12: ...e restored 6 2 When performing any adjustments and examining in the machine the power supply must be turned off 6 3 When installing or removing the indenter the mounting surface should be clean and dr...

Page 13: ...atic measurement Field proven under severe sample surface conditions 3 Single camera with four magnifications 1 3M pixel CMOS USB camera with two camera tubes with each tube height having two magnific...

Page 14: ...atic measurement Automatically or manually measures the indentation diameters on the calibration sample and the test sample and automatically interpolates the HB value for the test sample 4 Conversion...

Page 15: ...18 311 304 297 290 284 278 85 6 85 0 84 5 83 9 83 4 82 8 82 3 81 8 81 2 80 7 80 1 79 6 79 0 78 5 78 0 77 4 76 8 76 3 75 9 75 2 74 7 74 1 73 6 73 1 72 5 72 0 71 5 70 9 70 4 69 9 69 4 68 9 68 4 67 9 67...

Page 16: ...236 231 226 221 216 211 206 201 196 192 188 184 180 176 173 170 167 164 161 158 155 152 150 147 145 143 141 139 137 135 131 129 127 125 124 61 5 60 9 60 2 59 5 58 9 58 3 57 6 57 0 56 4 55 8 55 2 54 6...

Page 17: ...611 603 596 590 583 576 570 563 557 551 545 539 533 527 521 516 510 62 61 60 59 58 57 56 55 54 53 52 51 50 49 48 47 46 45 44 43 42 41 40 39 38 37 36 35 34 33 32 31 30 29 28 27 26 25 24 23 22 21 20 76...

Page 18: ...s Azcapotzalco C P 02020 Mexico Tel 01 55 3622 7000 Fax 01 55 5319 4000 BEIJING CHINA Phase II Measuring Instruments Beijing Ltd Room 301 Bldg 2 Qing Yuan Xi Li Haidian District Beijing 100192 China T...

Reviews: