Replacing the tube sensor (19-7815-02)

Dismounting

1.   Disconnect the two cables (8).

2.   Dismount the snap ring (7).

3.   Release the fl at cable and connector (1) from the arm.

4.   Remove the cover (3).

5.   Support the arm and the tube sensor from underneath 

and with a screwdriver or similar, press on top of 
sensor shaft (2) until it comes free from the snap 
ring (4).

6.   Pull the tube sensor out of the hole and unhook the 

spring (6).