GE Healthcare
Instructions 59-6144-00 AK
FRAC-100
Replacing the tube sensor (19-7815-02)
Dismounting
1. Disconnect the two cables (8).
2. Dismount the snap ring (7).
3. Release the fl at cable and connector (1) from the arm.
4. Remove the cover (3).
5. Support the arm and the tube sensor from underneath
and with a screwdriver or similar, press on top of
sensor shaft (2) until it comes free from the snap
ring (4).
6. Pull the tube sensor out of the hole and unhook the
spring (6).
Fig 1.
Fig 2.
Fig 3.