16
Connection "DeviceNet"
6.4 LED operation
NET
MOD
6.5 DeviceNet objects
Identity
Message Router
This object provides no attributes and services.
DeviceNet
Status
Meaning
Action
Off
Device not on bus
Supply operating voltage
Wait for address test (approx. 2 s)
Green flashing
On bus, no master allocated
Create connection with master
Green illuminated
On bus, master allocated
Red flashing
Connection to master expired
Test connection to master
Red illuminated
Bus error or duplicate device address
allocated
Check bus
Check device address
Check baud rate
Status
Meaning
Action
Off
No voltage supply
Supply operating voltage
Green illuminated
Device ready for operation
Red flashing
Device error
Resolve error
Path
Name
Data type
Service
Comment
1.0.1
Revision
UINT
get
1.0.2
Max Instance
UINT
get
1.0.3
Number of Instances
UINT
get
Instance 1
1.1.0
(Instance)
USINT
res, gaa
1.1.1
Vendor ID
UINT
get
527 (Pfeiffer Vacuum)
1.1.2
Device Type
UINT
get
33 (Turbomolecular Vacuum
Pump Device)
1.1.3
Product Code
UINT
get
5377
1.1.4
Revision
STRUCT of
get
Major Revision
USINT
Minor Revision
USINT
1.1.5
Status
WORD
get
1.1.6
Serial Number
UDINT
get
1.1.7
Product Name
SHORT_STRING get
TM 700 DN
1.1.100
Status Code
SHORT_STRING get
Path
Name
Data type
Service
Comment
3.0.1
Revision
UINT
get
Instance 1
3.1.0
(Instance)
all, re
3.1.1
MAC ID
USINT
get, (set) set only at switch setting "P"
3.1.2
Baud Rate
USINT
get, (set) set only at switch setting "P"
3.1.5
Allocation Information
STRUCT of
get
Allocation Choice Byte
BYTE
Master’s MAC ID
USINT
3.1.6
MAC ID Switch Changed
BOOL
get
3.1.7
Baud Rate Switch Changed
BOOL
get
3.1.8
MAC ID Switch Value
USINT
get
3.1.9
Baud rate Switch Value
USINT
get