Symptom / malfunction
messages
Possible causes
Remedy
Poor peak shape
(Poor peak shape)
Analyzer contaminated
● Degas the ion source.
● Bake out the analyzer.
● Replace the filament unit and/or the ion source.
● Replace the analyzer.
● Contact
.
System pressure too high
● Ensure that the pressure is below the maxi-
mum operating pressure: < 5 × 10
-4
hPa.
Ion source parameters faulty.
● Contact
.
Mass filter defective or contaminat-
ed
● Contact
.
Ion source or analyzer magnetized
● Rectify the cause of the magnetic fields.
Ion source or analyzer in close
proximity to magnetic fields
● Rectify the cause of the magnetic fields.
Mass calibration necessary
● Carry out a mass calibration.
Error in the electronic unit (QME)
● Contact
.
High noise level
(High noise level)
System grounding
● Make sure that the vacuum system is ground-
ed.
Scan speed is too quick for the am-
plification setting
● Increase the dwell time (Dwell).
● Contact
.
Electronic unit (QME) not attached
to the analyzer correctly
● Make sure that the electronic unit (QME) has
been completely pushed onto the analyzer .
● Ensure that the contacts are straight and are
not bent or damaged.
Electron multiplier defective or
worn
● Check the gain of the electron multiplier.
● Replace the analyzer.
● Contact
.
Error in the electronic unit (QME)
● Contact
.
Low sensitivity
● Refer to the operating and measuring error
"Low sensitivity".
Tbl. 17:
Operating and measuring errors
10.1.4 Calling software error messages from PV MassSpec
Symptom / malfunction messages
Possible causes
Remedy
PV MassSpec error
You can find error messages and remedies in the help files for PV MassSpec.
Tbl. 18:
Software error messages from PV MassSpec
Malfunctions