Excerpt of an exemplary a test report
View test report
1. Open your web browser.
2. In the address bar, enter:
http://<IP ADDRESS>//testreport.pdf
– In place of <IP ADDRESS>, enter the PrismaPro IP address.
The test report PDF file opens.
9.3 Maintaining the analyzer QMA 250
NOTICE
Filament damage from incorrect handling
Filaments are highly sensitive, pre-assembled units. You can easily damage filaments by incorrect
handling.
► Allow only qualified personnel to replace filaments.
► Never touch the filament, even with gloves.
► Never bend the filament unit.
► Never tilt the filament unit, as otherwise you will bend the formation room.
Maintenance
44/86