Exhaust pressure of the vacuum pump
≤ 1500 hPa absolute
Exhaust pressure at OME
max. atmospheric pressure
Tbl. 3:
Limits of use of the product
2.6 Proper use
► Use the vacuum pump for vacuum generation only.
► When pumping media with an oxygen concentration level of ≥ 21 %, only use perfluorinated, syn-
thetic oils (F4, F5, A113) as operating fluid.
► Adhere to the installation, commissioning, operating, and maintenance instructions.
► Do not use any accessory parts other than those recommended by Pfeiffer Vacuum.
2.7 Foreseeable improper use
Improper use of the product invalidates all warranty and liability claims. Any use that is counter to the
purpose of the product, whether intentional or unintentional, is regarded as misuse, in particular:
● Pumping of corrosive media (exception: C version of the rotary vane vacuum pumps)
● Pumping radioactive media
● Pumping of gases that introduce an ignition source to the suction chamber
● Pumping of gases that contain impurities such as particles, dust, or condensate
● Pumping explosive media
● Pumping of media with a propensity to sublimation
● Pumping of fluids
● Use of the vacuum pump in potentially explosive atmospheres
● Use of the vacuum pump outside the specified area of application
● Use for pressure generation
● Use in strong electrical, magnetic, or electromagnetic fields
● Connection to vacuum pumps or equipment which are not suitable for this purpose according to
their operating instructions
● Connection to devices with exposed live parts
● Use of accessories or spare parts not listed in these instructions
● Use of operating fluids other than those specified by Pfeiffer Vacuum
● Use of D1 or mineral oil as operating fluid with an oxygen concentration level of > 21 %
Mineral oils are combustible and ignite in high temperatures and when they come into contact with
pure oxygen. This oils oxidize heavily and thus lose their lubricating capacity.
2.8 Responsibilities and warranty
Pfeiffer Vacuum shall assume no responsibilities and warranty if the operating company or a third party:
● disregards this document
● does not use the product for its intended purpose
● carries out any modifications to the product (conversions, changes, etc.) that are not listed in the
corresponding product documentation
● operates the product with accessories that are not listed in the corresponding product documenta-
tion
The operator is responsible for the process media used.
2.9 Owner requirements
Safety-conscious working
1. Only operate the product in a technically flawless state.
2. Operate the product in line with its intended purpose, safety and hazard-conscious and only in
compliance with these operating instructions.
3. Fulfill the following instructions and monitor the observation of the following instructions:
– Proper use
– Generally applicable safety instructions and accident prevention regulations
– International, national and locally applicable standards and guidelines
– Additional product-related guidelines and regulations
Safety
15/60