7 Parameter set
7.1 General
All variables of the vacuum pump relevant for the function are stored as parameters in the electronic
drive unit. Each parameter has a three-digit number and a description. Parameters can be used via
Pfeiffer Vacuum display and control panels or via RS-485 with the Pfeiffer Vacuum protocol.
Conventions
Parameters are printed in bold as
three-digit numbers
in
square brackets
. The designation can also
be specified.
Example:
[P:312]
Software version
Factory settings
The electronic drive unit is pre-programmed in the factory. This makes a more direct and safe operation
of the vacuum pump possible without additional configuration.
You can configure the following parameters for the specific process:
●
[P:026]
: rotation speed setting mode on/off.
●
[P:707]
: specification of the rotation speed setting mode with 30 to 170 % of the nominal rotation
speed.
●
[P:002]
: stand-by mode on/off.
#
Three digit number of the parameter
Display
Display of parameter description
Description
Brief description of the parameters
Functions
Function description of the parameters
Data type
Type of formatting of the parameter for the use with the Pfeiffer Vacuum protocol
Access type
R (read): Read access; W (write): Write access
Unit
Physical unit of the described variable
min. / max.
Permissible limit values for the entry of a value
default
Factory default pre-setting (partially pump-specific)
The parameter can be saved persistently in the electronic drive unit
Tbl. 7:
Explanation and meaning of the parameters
7.2 Control commands
#
Display
Designations
Functions
Data
type
Ac-
cess
type
Unit min. max. de-
fault
can
be
stor-
ed
002 Standby
Standby
0 = off
1 = on
0
RW
0
1
0
x
009 ErrorAckn
Malfunction ac-
knowledgement
0
W
1
1
010 PumpgStatn
Pump
0 = off
1 = on
0
RW
0
1
0
x
019 Cfg DO2
Configuration
output DO2
1 = No error
2 = Error
5 = Set rotation speed
reached
6 = Pump on
9 = "0"
10 = "1"
11 = Remote priority
active
7
RW
0
21
5
x
Parameter set
29/52