Decommissioning
29
8
Decommissioning
8.1
Shutting down for longer periods
If the turbopump should be shut down for longer than a year:
Remove the vacuum pump from the system, if necessary.
Change the operating fluid reservoir.
Close the high vacuum flange of the turbopump.
Evacuate turbopump via the fore-vacuum flange.
Vent turbopump via the venting connection with oil-free, dry air or inert gas.
Close the flange openings by using the original protective covers.
Close further connection ports by using the corresponding protective covers.
Place pump upright on rubber feet.
Store the pump only indoors at temperatures between -25 °C and +55 °C.
In rooms with moist or aggressive atmospheres, the pump must be airproof shrink-
wrapped in a plastic bag together with a bag of desiccant.
8.2
Re-starting
Check turbopump for contamination and moisture.
Clean the turbopump externally with a lint-free cloth and little industrial alcohol.
If necessary, have Pfeiffer Vacuum Service clean the turbopump completely.
If necessary, have the bearings replaced. Take into account the total running time.
Change the operating fluid reservoir.
Installation and commissioning in accordance with the operating instructions.
8.3
Disposal
Products or parts thereof (mechanical and electrical components, operating fluids, etc.)
may cause environmental burden.
Safely dispose of the materials according to the locally applicable regulations.
WARNING
Contamination of parts and operating fluid by pumped media is possible.
Poisoning hazard through contact with materials that damage health.
In the case of contamination, carry out appropriate safety precautions in order to pre-
vent danger to health through dangerous substances.
Decontaminate affected parts before carrying out maintenance work.
NOTICE
Restarting
The serviceability of the operating fluid of the turbopump without operation is a maxi-
mum of 4 years. Before restarting after a shut-down of
4 years or longer
, carry out the
following work:
Replace the operating fluid reservoir
Replace bearings
Follow the maintenance instructions and inform Pfeiffer Vacuum