General safety precautions
► Do not expose body parts to the vacuum.
► Observe the safety and accident prevention regulations, if necessary wear personal protective
equipment.
► Check all safety measures at regular intervals.
► Always ensure a secure connection to the earthed conductor (PE), protection class I.
► During operation, make sure that plug-and-socket connections are securely fitted.
► Never operate the vacuum pump with open vacuum flange.
► Never make your own conversions or modifications to the vacuum pump.
► Before returning the vacuum pump, observe the notes in the chapter Service.
2.4 Limits of use
Parameter
HiLobe
Installation location
weatherproof (internal space)
Installation altitude
max. 2000 m above sea level
1)
Installation surface flatness
horizontal, max. permissible angle of inclina-
tion: ± 3°
Ambient temperature
+5
℃
to +40
℃
Relative humidity of air
max. 85 %
Permanent intake pressure in circulation mode < 1100 hPa (abs.)
Permanent intake pressure in vacuum mode
depending on max. pressure differential
Max. gas temperature, pressure side
250 °C
Loss of pumping speed for backing pump
≤ 10:1
Degree of pollution
2
Protection class
I
Overvoltage category
II
Permissible protection degree
IP54
Type 12 according to UL 50E
Tbl. 3:
Permissible ambient conditions
20
10
0
30
40
50
60
n
[1/min]
3000
5000
9000
11000
7000
[
h
Pa
]
Δ
p
ma
x
Fig. 3:
Max. permissible pressure differential between input and output sides, in accordance
with pump rotation speed
1)
at installation altitudes > 1,000m above m.s.l. and an ambient temperature of 40 °C the rated power of the
motor reduces by around 10 %
Safety