8.3.1 Draining the operating fluid
1
2
3
4
5
6
Fig. 10:
Draining the operating fluid
1 Filler screw
4
O-ring
2 O-ring
5
Drain screw
3 Sight glass
6
Collection receptacle
Draining the operating fluid
1. Unscrew the filler screw.
– Be careful with the O-ring.
2. Place a collection receptacle below the drain hole.
3. Unscrew the drain screw.
– Be careful with the O-ring.
4. Allow operating fluid to drain into collection receptacle.
8.3.2 Filling the operating fluid
1
2
3
Fig. 11:
Filling the operating fluid
1 Filler screw
3
Sight glass
2 O-ring
Maintenance
37/52