Scan this QR code or click 

here

 and view a document that will help you determine the de-

gree of aging of the P3 operating fluid in clean processes using a color chart (according to 
DIN 51578). The document with the number PK0219 can also be obtained from the 

Pfeiff-

er Vacuum Download Center

 on request.

Prerequisites

● Vacuum pump switched off
● Vacuum pump is vented to atmospheric pressure on the suction side
● Vacuum pump cooled

Required aids

● Test tube
● Pipette with flexible hose

Determine degree of aging of P3 operating fluid

1. Unscrew the filler screw.
2. Use the pipette to extract a sample of the operating fluid from the filling port.
3. Pour the sample into a test tube.
4. Check the sample in bright light.
5. If it is a reddish-brown color at the latest (corresponding with color identification number 5), 

change the operating fluid.

6. Screw in the filler screw.

7.3.2 Draining the operating fluid

 WARNING

Health hazard and risk of environmental damage from toxic contaminated operating fluid

Toxic process media can cause operating fluid contamination. When changing the operating fluid, 
there is a health hazard due to contact with poisonous substances. Illegal disposal of toxic substan-
ces causes environmental damage.

► Wear suitable personal protective equipment when handling these media.
► Dispose of the operating fluid according to locally applicable regulations.

 CAUTION

Scalding from hot operating fluid

Danger of burns when draining operating fluid if it comes into contact with the skin.

► Wear protective equipment.
► Use a suitable collection receptacle.

Cleaning by changing the operating fluid

Pfeiffer Vacuum recommends, in cases of heavy contamination with process residues, 
cleaning the inside of the vacuum pump with several operating fluid changes.

Prerequisites

● Vacuum pump switched off
● Vacuum pump vented to atmospheric pressure
● Vacuum pump has cooled so that it can be touched
● Operating fluid still warm

Maintenance

40/64

Summary of Contents for HENA 202

Page 1: ...OPERATING INSTRUCTIONS EN Translation of the Original HENA 202 302 Rotary vane pump...

Page 2: ...eiffer vacuum de Further operating instructions from Pfeiffer Vacuum can be found in the Download Center on our website Disclaimer of liability These operating instructions describe all models and var...

Page 3: ...duct features 18 3 4 Scope of delivery 19 4 Transportation and Storage 20 4 1 Transporting the vacuum pump 20 4 2 Storing the vacuum pump 21 5 Installation 22 5 1 Installing the vacuum pump 22 5 2 Con...

Page 4: ...45 7 6 Changing the gas ballast filter 46 7 7 Cleaning the oil air heat exchanger 47 7 8 Changing the operating fluid type 48 8 Decommissioning 49 8 1 Shutting down for longer periods 49 8 2 Recommiss...

Page 5: ...ings for Asian motor Hena 202 29 Tbl 8 Motor protection switch settings for global motor Hena 302 30 Tbl 9 Motor protection switch settings for Asian motor Hena 302 30 Tbl 10 Maintenance intervals 38...

Page 6: ...12 Double star circuit 28 Fig 13 Delta connection 28 Fig 14 Star circuit 28 Fig 15 Connection example with PTC thermistor tripping unit 30 Fig 16 Opening and closing the gas ballast valve 34 Fig 17 Re...

Page 7: ...cuum Download Center 1 1 2 Variants These instructions apply to HenaLine vacuum pumps Pump type Pump version Hena 202 Standard version of rotary vane pump Hena 302 Standard version of rotary vane pump...

Page 8: ...D04 141 111 Rating plate example Rating plate of the rotary vane pump Motor rating plate not shown Warning hot surface This sticker warns of injuries caused by high temperatures as a result of touchi...

Page 9: ...y class ID Internal diameter N N Mean sea level PE Earthed conductor protective earth PTC Temperature dependent resistor positive temperature coefficient WAF Width across flats RSSR Radial shaft seal...

Page 10: ...rty Notes tips or examples indicate important information about the product or about this docu ment 2 2 Safety instructions All safety instructions in this document are based on the results of the ris...

Page 11: ...cuum pump not attached tipping over due to changes in the center of gravity or incorrect loading Serious injuries due to trapping or crushing of limbs e g feet are the re sult Do not use the vacuum pu...

Page 12: ...as a climbing aid Do not exert any force on the product Ensure that the product has a safe centre of gravity when mounting components Wear protective equipment e g protective shoes CAUTION Risk of inj...

Page 13: ...pose of the operating fluid according to locally applicable regulations WARNING Health hazard through poisoning from toxic contaminated components or devices Toxic process media result in contaminatio...

Page 14: ...age Take suitable safety precautions and prevent health hazards or environmental pollution by toxic process media Decontaminate affected parts before carrying out maintenance work Wear protective equi...

Page 15: ...ust pressure Atmospheric pressure Tbl 3 Limits of use of the product 2 5 Proper use Use the vacuum pump for vacuum generation only When pumping media with an oxygen concentration level of 21 only use...

Page 16: ...al ists are people responsible for construction mechanical installation troubleshooting and maintenance of the product and who have the following qualifications Qualification in the mechanical field i...

Page 17: ...volume of each chamber changes periodically as the rotor rotates This causes the gas to be continuously drawn in at the vac uum connection and compressed in the suction chamber by the rotation of the...

Page 18: ...y vane pump The gas ballast valve affects the final pressure of the rotary vane pump 3 2 Identifying the product To ensure clear identification of the product when communicating with Pfeiffer Vacuum a...

Page 19: ...3 4 Scope of delivery Rotary vane vacuum pump with motor Operating fluid Locking caps for vacuum and exhaust connection Operating instructions Product description 19 64...

Page 20: ...em resulting in the vane breaking when the vacuum pump is switched on Always transport the vacuum pump horizontally or without operating fluid filling Fill in the operating fluid only at the final ins...

Page 21: ...um pump on an adequately sized level surface 4 2 Storing the vacuum pump Storage Pfeiffer Vacuum recommends storing the products in their original transport packaging Procedure 1 Seal the vacuum and e...

Page 22: ...bserve the permissible angle of inclination of 1 3 Screw the vacuum pump onto the rubber metal buffers on the mounting surface 4 When installing the pump in a closed housing ensure adequate air circul...

Page 23: ...f injury from bursting as a result of high pressure in the exhaust line Faulty or inadequate exhaust pipes lead to dangerous situations e g increased exhaust pressure There is a danger of bursting Inj...

Page 24: ...er consultation with Pfeiffer Vacuum Approved operating fluid P3 standard operating fluid D2 for special applications such as higher operating temperatures The operating fluid type is listed on the ra...

Page 25: ...r NOTICE Vacuum pump damage caused by different cooling characteristics For applications with short switching cycles or increased ambient temperature there is a risk of dam aging the rotor after switc...

Page 26: ...hat electrical installations are only carried out by qualified electricians Provide adequate grounding for the device After connection work carry out an earthed conductor check CAUTION Danger of injur...

Page 27: ...s marked with the symbol The voltage between the incoming mains supply lines is called mains voltage The mains current is the current flowing in the incoming supply lines Connect the three phase motor...

Page 28: ...the connection diagram 5 6 3 Connect three phase motor with 12 pin terminal board There are 3 different circuit configurations Double star circuit for low voltage Delta connection for medium voltage...

Page 29: ...ted current over a 2 minute period is permissible for the drive motors in accordance with EN 60034 1 without tripping the motor protection switch Procedure 1 Take the settings for the motor protection...

Page 30: ...up time 30 s Shutdown can take place immediately 5 6 7 Connecting the PTC thermistor tripping unit Tripping units store the shut down Pfeiffer Vacuum recommends connecting motors with PTC in the stat...

Page 31: ...Procedure After shut down switch the tripping unit back on manually via the installed RESET button or via the external RESET S3 Switching on mains detected as automatic RESET Installation 31 64...

Page 32: ...r of injury from bursting as a result of high pressure in the exhaust line Faulty or inadequate exhaust pipes lead to dangerous situations e g increased exhaust pressure There is a danger of bursting...

Page 33: ...ation Cycle operation with maximum 10 cycles per hour is possible Longer operating phases and short downtimes permit a functionally safe operating condi tion of the vacuum pump Operating conditions Th...

Page 34: ...gas ballast valve supplies air to the working chamber of the vacuum pump periodically at the begin ning of the compression phase When pumping down vapors this air prevents condensation within cer tai...

Page 35: ...ing Install an additional shut off valve in the intake line Shut off the intake line immediately after switching off the vacuum pump NOTICE Contamination of the vacuum chamber from operating fluid ba...

Page 36: ...edure 1 If required switch the vacuum pump off in each pressure range 2 Switch the mains switch off or securely disconnect the drive motor from the mains 3 Shut off the cooling water flow Operation 36...

Page 37: ...Wear protective equipment e g protective shoes CAUTION Danger of injury from moving parts After a power failure or a standstill as a result of overheating the motor restarts automatically There is a r...

Page 38: ...tenance work of Mainte nance Level 1 and Maintenance Level 3 revision If the required intervals listed below are exceeded or if maintenance work is carried out improperly no warranty or liability clai...

Page 39: ...amentally when filling refilling or changing operating fluid you must always use the operating fluid type specified on the rating plate Should process conditions change you can convert to a different...

Page 40: ...Draining the operating fluid WARNING Health hazard and risk of environmental damage from toxic contaminated operating fluid Toxic process media can cause operating fluid contamination When changing t...

Page 41: ...lace a collection receptacle below the drain hole 3 Unscrew the drain screw Be careful with the O ring 4 Fully drain the operating fluid 5 Seal the drain hole 7 3 3 Filling up operating fluid Cleaning...

Page 42: ...level 1 Required aids Collection receptacle 7 l Change operating fluid for cleaning 1 Operate the vacuum pump with the gas ballast open until it is warm 2 Perform an operating fluid change 3 Check the...

Page 43: ...ter in the operating fluid separator WARNING Health hazard through poisoning from toxic contaminated components or devices Toxic process media result in contamination of devices or parts of them Durin...

Page 44: ...10 2 3 1 Fig 21 Dismantling the exhaust filters 1 Operating fluid separator 3 Exhaust filter 2 Screw 4x Procedure 1 Remove the exhaust line 2 Remove the 4 screws and remove the exhaust connection 3 Pu...

Page 45: ...screws Tightening torque 6 Nm During operation the new exhaust filter becomes saturated with operating fluid It is therefore normal for the operating fluid fill level to drop slightly after the filter...

Page 46: ...arts 2 Examine all parts for wear Mounting the intake strainer 1 Mount the parts of the non return valve 2 Insert the intake strainer into the intake duct 3 Fit the vacuum connection 4 Screw on the cy...

Page 47: ...intenance work there is a risk to health from contact with these poisonous substances Illegal disposal of toxic sub stances causes environmental damage Take suitable safety precautions and prevent hea...

Page 48: ...O rings 2 Screw in the 2 screws Tightening torque 20 Nm 7 8 Changing the operating fluid type Possibilities for changing the operating fluid type The operating fluid type can be changed between minera...

Page 49: ...uum pump in dry dust free rooms within the specified ambient conditions 9 Pack the vacuum pump together with a drying agent in a plastic bag and seal the vacuum pump airtight if it is to be stored in...

Page 50: ...ponents must be disposed of in accordance with the applica ble regulations relating to environmental protection and human health with a view to reducing natural resource wastage and preventing polluti...

Page 51: ...he nameplate Should malfunctions occur you can find information about potential causes and how to fix them here Problem Possible causes Remedy Vacuum pump will not start up No mains voltage or volt ag...

Page 52: ...e that connections are short and cross sections ade quately dimensioned Exhaust pressure too high Check exhaust line outlet open ings and exhaust side accesso ries Loss of operating fluid Operating fl...

Page 53: ...ity You can find more detailed information and addresses on our homepage in the Pfeiffer Vacuum Service section You can obtain advice on the optimal solution for you from your Pfeiffer Vacuum represen...

Page 54: ...duct in suitable stable transport containers only e Maintain applicable transport conditions ERKL RUNG KONTAMINIERUNG 6 Attach the contamination declaration to the outside of the packag ing 7 Now send...

Page 55: ...ges 12 2 Operating fluids Maintenance level 1 Select the respective operating fluid type and quantity under consideration of the details on the rating plate Description Order number P3 mineral oil 1 l...

Page 56: ...he processes Magnetized gas ballast valve Solenoid valve for remote control of gas ballast PTC resistor tripping device Monitors the motor winding temperature 13 2 Ordering accessories Description Ord...

Page 57: ...87 Pa m3 s 10 1 592 7 5 9 87 sccm 1 69 10 2 1 69 10 3 1 1 27 10 2 1 67 10 2 Torr l s 1 33 0 133 78 9 1 1 32 atm cm3 s 1 01 0 101 59 8 0 76 1 Tbl 16 Conversion table Units for gas throughput 14 2 Techn...

Page 58: ...hPa Rated power 50 Hz 5 5 kW 5 5 kW Rated power 60 Hz 6 6 kW 6 6 kW Mains connection voltage range 5 5 Nominal pumping speed at 50 Hz 300 m h 300 m h Nominal pumping speed at 60 Hz 360 m h 360 m h Wat...

Page 59: ...170 410 2 50 4x 4x M10 5 55 305 1 5 512 2 108 248 397 2 G2 DIN ISO 228 1 G2 DIN ISO 228 1 G2 DIN ISO 228 1 153 325 25 2x DN19 37 Fig 26 Dimensions Hena 202 2x DN19 Only for Hena 202 with water cooling...

Page 60: ...4x 4x M10 201 G2 DIN ISO 228 1 108 248 350 25 20 535 2 325 397 2 G2 DIN ISO 228 1 G2 DIN ISO 228 1 32 2x DN19 37 1 5 1 5 1 5 Fig 27 Dimensions Hena 302 2x DN19 Only for Hena 302 with water cooling Te...

Page 61: ...stances in Electrical and Elec tronic Equipment Regulations 2012 Harmonized standards and applied national standards and specifications ISO 12100 2010 EN 60204 1 2018 ISO 13857 2019 EN IEC 61000 6 2 2...

Page 62: ...U Restriction of the use of certain hazardous substances delegated directive 2015 863 EU Harmonized standards and applied national standards and specifications DIN EN ISO 12100 2011 DIN EN 60204 1 201...

Page 63: ...63 64...

Page 64: ...PD0108 ed B Date 2210 P N PD0108BEN...

Reviews: