Parameter
Cooling water
Appearance
● filtered
● mechanically clear
● visually clear
● no turbidity
● no sediment
● free from grease and oil
pH value
7 to 9
Carbonate hardness, max.
10 °dH
12.53 °e
17.8 °fH
178 ppm CaC0
3
Chloride content, max.
100 mg/l
Sulfate content, max.
240 mg/l
Carbonic acid content, max.
not detectable
Ammonia content, max.
not detectable
Electrical conductivity, max.
500 µS/cm
Particle size, max.
150 µm
Cooling water temperature
see "Technical data"
Cooling water flow
see "Technical data"
Feed line overpressure, max.
6000 hPa
Tbl. 5:
Requirements on the cooling water composition
Required material
● Cooling water hose,
ID 19 mm
● Hose clamps
1
2
3
Fig. 7:
Cooling water connection
1 Cooling water outlet
3
Oil/water heat exchanger
2 Cooling water inlet
Making the cooling water connection
1. Connect the cooling water lines:
– Cooling water inlet
– Cooling water outlet
2. Ensure that the cooling water outlet is unpressurized.
3. Use a dirt trap in the inlet.
Installation