The operating fluid type is listed on the rating plate
► Please refer to rating plate for type and quantity of intended operating fluid.
max.
min.
1
2
3
4
Fig. 8:
Fill operating fluid
1 Filler screw with o-ring
3
Rating plate
2 Filler hole
4
Sight glass
Consumable
● Operating fluid of the vacuum pump
Required tools
● Allen key,
WAF 8
● Calibrated torque wrench (tightening factor ≤ 2.5)
Fill operating fluid
1. Unscrew the filler screw.
2. Refill the operating fluid:
– Initial filling with cold vacuum pump: A maximum of 3/4 of the min./max. range.
3. Screw in the filler screw again.
– Be careful with o-ring.
– Tightening torque:
3.0 Nm
5.5 Connect to mains power supply
DANGER
Danger to life from electric shock
Touching exposed and voltage-bearing elements causes an electric shock. Improper connection of
the mains supply leads to the risk of touchable live housing parts. There is a risk to life.
► Before the installation, check that the connection leads are voltage-free.
► Make sure that electrical installations are only carried out by qualified electricians.
► Provide adequate grounding for the device.
► After connection work, carry out an earthed conductor check.
CAUTION
Danger of injury from moving parts
After a power failure or a standstill as a result of overheating, the motor restarts automatically. There
is a risk of injury to fingers and hands if they enter the operating range of rotating parts.
► Safely disconnect motor from the mains.
► Secure the motor against reactivation.
► Dismantle the vacuum pump for inspection, away from the system if necessary.
Installation
24/60