5.9.2 Connection to external power supply
WARNING
Danger to life from electric shock in the event of a fault
In the event of a fault, devices connected to the mains may be live. There is a danger to life from
electric shock when making contact with live components.
► Always keep the mains connection freely accessible so you can disconnect it at any time.
The pump operates with a 48 VDC external power supply. The power supply and cable are available as
accessories (see chapter "Accessories").
It can also operate with an external power supply and a cable supplied by the customer with the re-
quired characteristics .
If the pump is connected to a main 48 VDC network (as defined by standard EN 61000-6), add a filter
and an overvoltage clipping system on the power line at the pump inlet: contact our service center. The
“-” pole of the 48 VDC is grounded internally.
1
2
4
3
5
6
7
Fig. 12:
48 VDC power supply male connector
1 Index
5
Not connected
2 Po 48 VDC
6
Pump power cable plug (Ref.: UTS6JC104S supplier: Souriau France)
3 Polarity - 48 VDC
7
Power cable
4 Ground potential
Pump power supply
1. Mark the main index on the pump’s male electrical connector.
2. Mark the main index on the female connector of the corresponding supply cable.
3. Insert the female connector in compliance with the indexes
and turn until you reach the locking
position
.
Air cooled version
► If the pump is equipped with a fan, connect it via the connector to the front panel of the external
power supply.
5
4
0V
48V
1
2
3
Fig. 13:
Fan wiring to 48 VDC external power supply (accessory)
1 Connector for fan power supply (air cooled
version)
4
Pump power cables: +++ (48 V), --- (0 V)
2 Connector for pump power supply
5
Fan power cables [ 1 or blue (-) (0 V), 2 or brown
(+) (48 VDC) ]
3 Mains connector
ATH 500 M version (only)
► Power the purge solenoid valve with 24 VDC (power supplied by the customer).
Installation
36/78