Repairs
36
Replacement of actuators
Depending upon the model of the Stander's seat,
the number of actuators can vary.
Removal
1.
With the seat in the sitting position, run the
seat lift up in order to access it more easily
from below.
2.
Set the Circuit breaker for the chassis to the
”OFF” position. For more information, see the
chassis service manual.
3.
Remove the seat plate in order to access the
seat function board. The seat plate is
mounted using four screws, see Fig. 85.
4.
Remove the protective cover sitting above the
actuator's electrical connections. The
protective cover is mounted using three
screws, see Fig. 86.
5.
Loosen the two terminals from the actuator,
see Fig. 87. Take note of its location for
purposes of later remounting.
Fig. 86. The electrical connections are located
behind a protective cover.
Fig. 87. The actuator's electrical connections.
Fig. 85. The seat plates are mounted
using four screws.