Repairs
Fig. 60. The sensor for the Stand-down stop
switch is mounted 3 mm under the other sensors.
4.
Remove the sensor by loosening its retaining
nuts, see Fig. 58.
5.
Loosen the cable connections, see Fig. 59.
Check the placement of the cables for
purposes of remounting later.
Fitting
Mounting is performed in the reverse order. Sensors
1-4 (see Fig. 56) are mounted at a height of 29 mm
from the frame side member, see Fig. 60. Sensor
number 5 (see Fig. 56) is mounted at a height of 26
mm from the frame side member, see Fig. 60. Other
sensors are mounted so that they protrude as far as
possible.
Fig. 58. The sensor is mounted with a nut.
Fig. 59. The sensors cable connection.
26 mm.