M400 Corpus HD
Troubleshooting - Diagnostics R-net LED control panel
7.2.7.6
Testing adjustment device
If the wheelchair is equipped with an adjustment device:
• Check that all motors move in the right direction.
• Make sure that the mechanical end stops are secured and that they
stop the adjustment device motors, and thus use the automatic end
stop tracking that is in the seat and light module (ISM).
7.2.7.7
Testing inhibit signal
Connect a suitable battery charger or equivalent inhibit connecting
device in the charging contact on the joystick module and check that
the wheelchair is prevented from running.
If inhibit contacts 2, 3, 4 and 5 are used for inhibiting or restricting
speed, suitable test must be performed in order to check that they are
functioning as intended.
7.3
Diagnostics R-net LED control panel
7.3.1
Battery voltage indicator
Each time the wheelchair is started, parts of its electronics are checked.
When a fault occurs in these parts, it is displayed on the control panel
battery voltage indicator and the indicator for speed or driving profile in
the form of one or more flashing LEDs.
Troubleshooting and repairs must always be performed by qualified
personnel with good knowledge of the wheelchair‘s electronics.
NOTICE
Error signals
Error messages are not displayed on the indicators while the wheelchair is
being driven. They appear when it is next started.
7.3.2
Steady
Everything is in order. The number of LEDs that light up depends on
the charge remaining in the batteries. If the batteries are fully charged,
all the LEDs light up.
7.3.3
Slowly flashing red LEDs, 1–2 LEDs
The batteries must be charged immediately.
7.3.4
Rapidly flashing, 1–10 LEDs
A fault has been detected in the wheelchair‘s electronics and the
wheelchair may not be driven.
1.
Switch off the wheelchair.
2.
Check that all visible cables and the cable to the control panel are
connected correctly.
3.
Switch the wheelchair on again. If the fault persists, count the
number of flashing LEDs and check for a possible cause and remedy
in the following table.
4.
Do not use the wheelchair until the problem has been remedied or
you have received other information from your service provider.
96
Summary of Contents for M400 Corpus HD
Page 1: ...Service manual American English M400 Corpus HD ...
Page 2: ......
Page 4: ...Text Produced and published by Permobil Edition 3 Date 2018 10 18 Order no 334809 eng US ...
Page 6: ......
Page 15: ...M400 Corpus HD Specifications Wiring diagram Figure 4 Cabling overview Chassis 2 2 15 ...