216 . Torion T-9 User’s Guide
understanding of what each specific setting is used for and what it does. Contact a PerkinElmer
service representative for assistance when making changes to the Mass Spectrometer.
MS Calibrations Tab
The MS Calibrations tab contains calibration data that are created in the
Performance Validation
.
The following image shows the MS Calibration tab.
If
Calibrate data
is checked then either the Manual Calibration Data or the Auto Calibration Data is
used when looking at the data in
Data Review
or on the instrument. If this box is not checked then
the data are viewed as uncalibrated data. This means that the data are shown in their raw bin index
location with no mass assignments. This can be useful for some troubleshooting operations.
•
Manual Calibration Data
are data that are either hand entered by the user or are copied
from the Auto Calibration Data and then modified.
•
Auto Calibration Data
are the data generated from the
Performance Validation
process. These data cannot be changed. If the computer is not connected to an instrument
this table will be blank.
Summary of Contents for TORION T-9
Page 1: ...TORION T 9 Gas Chromatograph Mass Spectrometer GC MS User s Guide GC MS...
Page 4: ...iv Torion T 9 User s Guide...
Page 10: ...6 Torion T 9 User s Guide...
Page 11: ...1 Introduction...
Page 15: ...2 Safety Practices...
Page 24: ...20 Torion T 9 User s Guide...
Page 31: ...3 System Overview...
Page 40: ...36 Torion T 9 User s Guide...
Page 41: ...4 Basic Custodion Operations...
Page 49: ...5 Basic Operation...
Page 116: ...112 Torion T 9 User s Guide...
Page 117: ...6 Software...
Page 257: ...7 Advanced Operations...
Page 260: ...256 Torion T 9 User s Guide...
Page 261: ...8 Maintenance...
Page 268: ...264 Torion T 9 User s Guide...
Page 269: ...9 Troubleshooting...