Maintenance . 115
Replacing the Microscope Fuse
The microscope is not voltage sensitive and will operate at voltages between 100–240 V and
at 50–60 Hz.
1.
Switch off the microscope and disconnect it from the power supply.
2.
Insert a screwdriver into the slot at the side of the fuse drawer; pull out and flip to one
side over the mains inlet.
The fuse may now be removed.
3.
Fit the replacement fuse into the fuse drawer.
You require a 2.0 A time-lag, 250 V fuse (part number 04970839).
4.
Refit the fuse drawer.