Getting Ready to Use the Spotlight 200 . 45
7.
Remove the funnel and wait two minutes.
The liquid nitrogen settles down and bubbling slows.
8.
When the nitrogen stops bubbling, refit the detector cap.
The filled dewar cools the MCT to the correct operating temperature for several hours.
After that, the dewar begins to return to room temperature, and the preamp switches
off power to the MCT.