76 . Spotlight 150 User's Guide
Adjusting the Height of the Crystal
CAUTION
To prevent serious damage to the crystal, do not move the Sample
Stage while the ATR crystal is in the lower, working position and
touching or near to the sample.
1.
Lower the manual ATR crystal onto the reference mirror by turning the bar clockwise
and slowly lowering the toggle bar.
You must always support both ends of the toggle bar when lowering or raising the
crystal.
OR
Click the
button to lower the automated ATR crystal onto the reference mirror.
2.
Using the knurled height-adjustment nut, adjust the height of the crystal so that it just
rests on the mirror under positive pressure (Figure 34).
Figure 34 Adjusting the Crystal Height (Manual ATR Objective)
Figure 35 Adjusting the Crystal Height (Automated ATR Objective)
Knurled height-adjustment nut