76 .
RamanMicro 200 Series Getting Started Guide
Fiber Optic Probe SOPs
NOTE: The SOPs included on the following pages are only appropriate for a standard fiber
optic probe. For details of how to acquire data using a triggered fiber optic probe, see
the
Raman Triggered Fiber Optic Probe
leaflet (L1321940).
SOP-7: Selecting the Fiber Optic Probe
1.
Ensure that the RamanMicro 200 is switched on and that the fiber optic probe is
connected to the instrument.
If not follow SOP-1.
2.
Select
Raman Instrument
from the Setup menu in Spectrum.
The Setup Instrument tabs are displayed in the Dialog pane.
3.
Select the Setup Instrument Basic tab.
Figure 40 Setup Instrument Basic tab
4.
Select the Fiber Optic Probe you want to use from the Accessory drop-down list.
The following warning message is displayed.