Removing the FOP
If it is necessary to remove the FOP, for example, for replacement or transportation, please
follow this procedure carefully:
1.
Ensure the attenuator cap is attached to the FOP.
2.
Switch off the instrument and remove the main power cable.
3.
Carefully move the spectrometer module to permit easier access to the rear panel.
4.
Remove the plug in the socket marked LED FIBER 1 (or LED FIBER 2 if you are
removing the second FOP from your RamanFlex 400F).
5.
Carefully remove the retained ends of the FOP by unscrewing and pulling the fiber optic
cables away from the couplers perpendicularly.
Be extremely careful when removing the FOP as you can easily damage the instrument.
6.
Replace the plastic caps on the ends of the FOP that have been removed from the rear
panel.
7.
Fasten the retained screw-in caps onto the couplers on the rear panel.
8.
Reconnect the power cables.
Summary of Contents for RamanFlex 400
Page 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 RamanFlex 400 Series Getting Started Guide ...
Page 35: ...Preparing your Facility ...
Page 40: ...40 RamanFlex 400 Series Getting Started Guide ...
Page 41: ...Installing your RamanFlex 400 ...
Page 61: ...Introduction to the RamanFlex 400 ...
Page 68: ...68 RamanFlex 400 Series Getting Started Guide ...
Page 69: ...Using the RamanFlex 400 ...