66 . Lambda 365 Users Guide
Tungsten Lamp Module Replacement
1.
Make sure that the main power is turned off. If the system and lamp(s) were turned
on at least 20 minutes should pass before the lamp changing process is started. This
time allows the lamps to cool and return to room temperature.
CAUTION
ATTENTION
Use the Tungsten Lamp Changing Tool (Part No. N4102048) when
replacing the Tungsten Lamp. In addition a pair of appropriate
disposable gloves should be worn during the lamp replacement
process. Please avoid touching the glass bulb section of the lamp this
can leave oil deposits on the bulb which can drastically reduce lamp
life.
Utilisez Tungsten Lamp outil Modification lors du remplacement de la
lampe de tungstène. En outre, une paire de gants jetables appropriés
doivent être portés pendant le processus de remplacement de la
lampe. S'il vous plaît éviter de toucher la section de l'ampoule de verre
de la lampe, cela peut laisser des dépôts d'huile sur le bulbe qui peut
réduire considérablement la vie de la lampe.
Tungsten Lamp Changing Tool (Part No. N4102048)
2.
Remove the white bolt covers (four each) using screwdriver and remove the M3*12
screws (four each) using a Phillips screwdriver to open the vent cover.
Summary of Contents for Lambda 365
Page 1: ...LAMBDA 365 Users Guide UV SPECTROMETERS ...
Page 4: ...4 Lambda 365 Users Guide ...
Page 5: ...Introduction 5 Introduction ...
Page 19: ...Safety Practices 19 Safety Practices ...
Page 30: ...30 Lambda 365 Users Guide ...
Page 31: ...Lambda 365 Installation 31 Lambda 365 Installation ...
Page 35: ...UV Express Software Installation 35 UV Express Software Installation ...
Page 37: ...Interface Setup 37 3 UV Express InstallShield Wizard starts Select Next ...
Page 41: ...Interface Setup 41 Interface Setup ...
Page 43: ...Interface Setup 43 5 Double click USBCom CDM_20828 6 Select Extract ...
Page 46: ...46 Lambda 365 Users Guide ...
Page 47: ...Setting USB Serial Ports 47 Setting USB Serial Ports ...
Page 48: ...48 Lambda 365 Users Guide Procedure 1 Select Computer Properties 2 Click on Device Manager ...
Page 53: ...Operating Procedure 53 Operating Procedure ...
Page 55: ...Accessories 55 Accessories ...
Page 62: ...62 Lambda 365 Users Guide ...