Software Reference
98
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Optics Normalization Tab (Continued)
Open Chip Door button
- Opens the front door to insert Test Chip C.
Warm and Scan/Scan/Abort button
:
•
Warm and Scan
-
Warms up the chip and laser, focuses the
optics, and measures the optical contrast.
•
Scan
- Scans without the warmup (available only after a scan is
complete or aborted, while laser and chip are still warmed up).
•
Abort
-
Stops the scan or warm currently in progress.
Apply button
- Saves the New Correction Factor to the
instrument’s non-volatile flash memory.
Plate Editor Tab
Use the Plate Editor tab to view, add, or change plate names and
measurements. To open the Plate Editor tab, touch the
Tools
button
on the
and then touch the
Plate Editor
button.
Figure 34. Plate Editor Tab
The following tabs display at the top of the Plate Editor tab:
•
Predefined Plates Tab
- Displays PerkinElmer-defined plates.
These plates are read-only and cannot be edited or deleted.
•
Custom Plates Tab
- Displays all user-created plates. Plates
can be added, deleted, and modified.